MV1E-5350V-B1
350 Watt
- Low noise!
- Complies with IEC/EN60601-1 – 3rd Edition
- Efficiency >80 %
The MV1E-5350V-B1 was designed for medical PC applications in a 1U chassis. With its output power of 350 W it is the perfect solution for image processing systems with powerful processors and graphic boards. It complies with the IEC/EN60601-1 and UL60601-1 standards. The new circuit design is distinguished by a very high efficiency of far more than 80 % over the complete range of performance. Thus, minimal noise development is achieved by a temperature-controlled fan regulation.
| Technical data
|
|
| Input voltage
|
90…264 VAC, active PFC
|
| Input frequency
|
47…63 Hz
|
| Input current
|
6 A (115 VAC) / 3 A (230 VAC)
|
| Inrush current
|
15 A (115 VAC) / 30 A (230 VAC)
|
| Efficiency
|
81–85 %
|
| Hold up time
|
>16 msec (115 VAC), 24 msec (230 VAC), full load
|
| Power-Good-Signal
|
Switch on delay 100…500 msec
Switch off delay 1 msec
|
| Protection
|
Short circuit protection: At each output, switch off / +5 Vsb, auto-recovery
Overload protection: 110…160 %, Switch off
Overvoltage protection: +3.3 V (+3.6…+4.3 V), +5 V (+5.6…+6.5 V), +12 V (+13.2…+15 V)
|
| Insulation voltage
|
Input / Chassis 1500 VAC
Input / Output 3000 VAC
|
| Earth leakage current
|
<200 µA /240 VAC
|
| Safety / EMC
|
TÜV (IEC/EN60601-1 3rd Edition), UL (UL60601-1), CE
|
| Operating temperature
|
-10…+50 °C
|
| Derating
|
From +20…+50 °C, 1 W / K (+3,3 V, +5 V, +12 V, +5 Vsb)
From +20…+50 °C, 0,01 A / K (–12 V)
|
| Storage temperature
|
-40…+80 °C
|
| Operating humidity
|
20…80 % RH, non-condensing
|
| Dimensions (WxDxH)
|
100 x 260 x 40.5 mm ±0.5 mm
|
| Weight (net)
|
1.5 kg
|
|
Article
No.
|
Output
voltage
|
Output current
min
|
Output current
max
|
Load
regulation
|
Ripple &
noise
|
| MV1E-5350V-B1
|
+3.3 V
|
0 A
|
20 A
|
±5 %
|
50 mV
|
| +5 V
|
0 A
|
20 A
|
±5 %
|
50 mV
|
| +12 V
|
0 A
|
25 A
|
±5 %
|
120 mV
|
| -12 V
|
0 A
|
1 A
|
+13 / -7 %
|
120 mV
|
| +5 Vsb
|
0.1 A
|
2 A
|
±5 %
|
50 mV
|
Max. output power is 350 W. Combined output load at +3.3 V and +5 V must not exceed 130 W, combined output load at +3.3 V, +5 V, and +12 V must not exceed 340 W. Ripple and noise was measured by a 20 MHz bandwidth limited oscilloscope with connected 100 µF electrolytic capacitor and 0.1 µF ceramic capacitor at each output. During a cross regulation test at -12V we recommend to keep the channel with higher output load at
80 % of its max. power and the channel with lower output load at 20 % of its max. power.
As a power component this PSU is for assembly purposes only and must not be operated in unassembled condition. The final assembly has to comply with the valid EMC and safety standards.
Technical documentation:
|