MPSM-5600V
600 Watt
- Ideal for medical image processing
- Complies with IEC/EN60601-1 – 3rd Edition
- High efficiency up to 85 %
The new PC power supply MPSM-5600V is designed for medical applications. At 600 watts output it is ideally suited for image processing systems equipped with a powerful graphics card. Its modern switching design is dimensioned for maximum energy savings. Even in partial tipping operation its efficiency will not drop below 82 %. High-quality parts ensure long availability of the powered system.
| Technical data
|
|
| Input voltage
|
90…264 VAC, active PFC
|
| Input frequency
|
47…63 Hz
|
| Input current
|
10 A (115 VAC) / 5 A (230 VAC)
|
| Inrush current
|
20 A (115 VAC) / 40 A (230 VAC)
|
| Efficiency
|
82…85 %
|
| Standby consumption
|
<1,6 W
|
| Hold up time
|
>16 msec
|
| Power-Good-Signal
|
Switch on delay 100…500 msec
Switch off delay 1 msec
|
| Protection
|
Short circuit protection: At each output, switch off / +5 Vsb, auto-recovery
Overload protection: 120…160 %, switch off
Overvoltage protection: +3.3 V (+3.6…+4.3 V), +5 V (+5.6…+6.5 V), +12 V (+13.2…+15 V)
|
| Insultaion voltage
|
Input / chassis 3100 VDC
Input / output 4242 VDC
|
| Earth leakage current
|
<300 µA, 115 VAC/230 VAC
|
| Safety / EMC
|
TÜV (IEC/EN60601-1 3rd Edition), UL (UL60601-1), CAN/CSA C22.2 No. 601.1-M90, CE
|
| Operating temperature
|
0…+40 °C
|
| Storage temperature
|
-40…+80 °C
|
| Operating humidity
|
20…80 % RH, non-condensing
|
| Dimensions (WxDxH)
|
150 x 160 x 86 mm ±0.5 mm
|
| Weight (net)
|
2.4 kg
|
|
Article
No.
|
Output
voltage
|
Output current
min
|
Output current
max
|
Load
regulation
|
Ripple &
Noise
|
| MPSM-5600V
|
+3.3 V
|
0.5 A
|
25 A
|
±5 %
|
50 mV
|
| +5 V
|
0.5 A
|
25 A
|
±5 %
|
50 mV
|
| +12 V
|
2 A
|
45 A
|
±5 %
|
120 mV
|
| -12 V
|
0 A
|
0.8 A
|
±5 %
|
150 mV
|
| +5 Vsb
|
0.1 A
|
3.5 A
|
±5 %
|
50 mV
|
Max. output power is 600 W, combined output power at +3.3 V and +5 V must not exceed 40 A. Ripple and Noise was measured by a 20 MHz bandwidth limited oscilloscope with connected 220 µF electrolytic capacitor and 0.1 µF ceramic capacitor at each output.During a cross regulation test we recommend to keep the channel with higher output load at 80 % of its max. power and the channel with lower output load at 20 % of its max. power.
As a power component this PSU is for assembly purposes only and must not be operated in unassembled condition. The final assembly has to comply with the valid EMC and safety standards.
Technical documentation:
|