MP1S-6400V-B1

400 Watt

  • Additional +24 VDC output
  • Ideal for applications with large monitors
  • High efficiency of up to 84 %

The MP1S-6400V-B1 is a medically approved 1U ATX PC power supply with an additional +24 VDC output. This allows the direct supply of medical peripherals including the areas of diagnostics, therapy and imaging with +24 VDC. The simultaneous supply of motherboard and peripherals with a single power supply also reduces the total earth leakage current of the medical system.

Technical data

 

Input voltage

100…240 V AC, active PFC

Input frequency

47…63 Hz

Input current

6 A (115 V AC) / 3 A (230 V AC)

Inrush current

70 A (115 V AC / 25°C) / 140 A (230 V AC / 25°C), cold start

Efficiency

84 % (at nominal load and 230 V AC)

Hold up time

>16 ms

Power-Good-Signal

Switch on delay 100…500 ms
Switch off delay 1 ms

Protection

Short circuit protection: +5 V, +12 V, switch off /+3.3 V, -12 V, +5 Vsb, auto recovery
Overload protection: 110…160 %, switch off
Overvoltage protection: +3.3 V (3.6…4.3 V), +5 V (+5.6…+6.5 V), +12 V (+13.2…+15 V)
Overcurrent protection: +3.3 V (11…16 A), +5 V (13.2…19.2 A), +12 V (16…24 A)

Earth leakage current

<0.3 mA, 115 V AC / 230 V AC

Safety / EMC

TÜV (IEC/EN60601-1 3rd Edition), UL (UL60601-1), CB, CCC, CE

Operating temperature

0…+40 °C

MTBF

149 868 h according to MIL-HDBK-217F at +25 °C with fan

Storage temperature

-20…+80 °C

Operating humidity

20…80 % RH, non-condensing

Dimensions (WxDxH)

81.5 x 250 x 40.3 mm ±0.5 mm

Weight (net)

1.2 kg

Article
No.

Output
voltage

Output current
min

Output current
max

Load
regulation

Ripple &
Noise

MP1S-6400V-B1

+24 V

0.1 A

6 A

±5 %

240 mV

+3.3 V

0.5 A

10 A

±5 %

60 mV

+5 V

1 A

12 A

±5 %

60 mV

+12 V

2 A

16 A

±5 %

120 mV

-12 V

0 A

0.3 A

±10 %

120 mV

+5 Vsb

0.1 A

2.5 A

±5 %

60 mV

Max. output power is 400 W. Power at +3.3 V, +5 V, +12 V, –12 V and +5 Vsb is max. 260 W, combined at +3.3 V and +5 V it is 90 W. Ripple and Noise was measured by a 20 MHz bandwidth limited oscilloscope with connected 100 μF electrolytic capacitor and 0.1 μF ceramic capacitor at each output.
As a power component this PSU is for assembly purposes only and must not be operated in unassembled condition. The final assembly has to comply with the valid EMC and safety standards.

Technical documentation:
© 2008-2012  |  BICKER ELEKTRONIK GmbH - Ludwig-Auer-Straße 23 - 86609 Donauwörth (DE)
Phone +49 (0)906 / 70595-0 - Fax +49 (0)906 / 70595-55