SNP-Z10M series
130 Watt
- 100 Watt continuous fanless power
- High efficiency up to 85 %
- EN60601-1 and UL60601-1 certified
The SNP-Z10M power supply series has been tested and approved for medical applications. By its high efficiency ranging from 80 to 85 % fanless and, thus, very quiet operation is made possible. The SNP-Z10M series is made up of high-grade components and provides overload, short circuit and overvoltage protection. Optionally, AC and DC cable harnesses as well as a casing are available for all models.
| Technical data
|
|
| Input voltage
|
90…264 VAC, 130…370 VDC
|
| Input frequency
|
47…63 Hz
|
| Input current
|
2.6 A / 115 VAC, 1.3 A / 230 VAC
|
| Inrush current
|
<30 A / 115 VAC, <60 A / 230 VAC
|
| Efficiency
|
80…85 %, depending on model
|
| Hold up time
|
App. 20 ms / 115 VAC and 100 Watt
|
| Protection
|
Short circuit protection: Switch off with auto-recovery
Overload protection: Switch off with auto-recovery
Overvoltage protection: Switch off (model / response threshold)
SNP-Z101M, -Z104M, -Z106M / (5.7 V…7 V), SNP-Z107M / (13.1 V…15.1 V), SNP-Z108M / (16.6 V…20.1 V), SNP-Z109M / (26.2 V…31 V), SNP-Z10TM / (55 V…62 V), SNP-Z10HM / (30 V…34 V)
|
| Line regulation
|
±0.5 % at rated load and input voltage change from 90 to 264 VAC
|
| Load regulation
|
±1 % (A1, load step ±40 % related to 60 % load, without fan)
±5 % (A2-A3, at 60 % load)
|
| Earth leakage current
|
<300 µA max. (240 VAC / 60 Hz)
|
| Safety / EMC
|
TÜV (EN60601-1), UL (UL60601-1), CE
|
| Operating temperature
|
0…+50 °C / 100 W with convection, SNP-Z106M: 0…+40 °C / 100 W
|
| Derating
|
+50…+70 °C / 2.5 % / K, SNP-Z106M: +40…+70 °C / 2.5 % / K
|
| Storage temperature
|
-20…+85 °C
|
| Operating humidity
|
5…95% RH, non-condensing
|
| Dimensions
|
127 x 76.2 x 32.4 mm ±0.5 mm / SNP-Z106M, -Z107M, -Z109M, -Z10HM
127 x 76.2 x 34.4 mm ±0.5 mm / SNP-Z101M, -Z104M, -Z108M, -Z10TM
|
| Weight (net)
|
0.3 kg
|
| Remote sense
|
Regulation range 0.5 V (except SNP-Z101M, SNP-Z104M)
|
|
Article
No.
|
Output
|
Output
voltage
|
Output current
min
|
Output current
rated load
|
Output current
max load
|
Ripple &
noise
|
| SNP-Z106M
|
A1
|
+5 V
|
0 A
|
20 A
|
26 A
|
<50 mVss
|
| SNP-Z107M
|
A1
|
+12 V
|
0 A
|
9 A
|
11 A
|
<100 mVss
|
| SNP-Z108M
|
A1
|
+15 V
|
0 A
|
7 A
|
8.7 A
|
<100 mVss
|
| SNP-Z109M
|
A1
|
+24 V
|
0 A
|
4.5 A
|
5.4 A
|
<100 mVss
|
| SNP-Z10HM
|
A1
|
+28 V
|
0 A
|
3.8 A
|
4.65 A
|
<100 mVss
|
| SNP-Z10TM
|
A1
|
+48 V
|
0 A
|
2.1 A
|
2.7 A
|
<150 mVss
|
| SNP-Z101M
|
A1
A2
A3
|
+5 V
+12 V
-12 V
|
0.5 A
0.3 A
0 A
|
11.5 A
3 A
0.8 A
|
15 A
4.1 A
1.1 A
|
<50 mVss
<120 mVss
<120 mVss
|
| SNP-Z104M
|
A1
A2
A3
|
+5 V
+15 V
-15 V
|
0.5 A
0.3 A
0 A
|
10 A
2.5 A
0.8 A
|
15A
4.1 A
1.1 A
|
<50 mVss
<150 mVss
<150 mVss
|
100 W continuous power without fan (SNP-Z106M 100 W / +40 °C, 80 W / +50 °C). For fan cooling at 130 W output power an 18CFM (30 m3 / h) fan is required, at 120 W a 10CFM (17 m3 / h) fan is required. Ripple and noise was measured by a 15 MHz bandwidth limited oscilloscope with connected 47 µF electrolytic and 0.47 µF ceramic capacitor at each output.
As a power component this PSU is for assembly purposes only and must not be operated in unassembled condition. The final assembly has to comply with the valid EMC and safety standards.