BEP-506M
80 Watt
- 60 Watt continuous fanless power!
- Operating temperature up to +70 °C
- Complies with IEC/EN60601-1 – 3rd Edition
The new PC power supply BEP-506M is very compact and was designed for low power ATX and mini ITX mainboards for medical uses. As it operates without active cooling it is the ideal power supply for fanless working applications in medical equipment. The BEP-506M is approved to the EN60601-1 by TÜV and meets the requirements of UL60601-1. For easy connection cable harnesses for input and output are available.
| Technical data |
|
|---|---|
| Input voltage |
90…264 VAC |
| Input frequency |
47…63 Hz |
| Input current |
2 A (115 VAC) / 1 A (230 VAC) |
| Inrush current |
30 A (115 VAC) / 60 A (230 VAC) |
| Efficiency |
≥75 % |
| Hold up time |
>20 msec at 230 VAC and rated load |
| Power-Good-Signal |
Switch on delay 100…500 msec |
| Protection |
Short circuit protection: at each output, switch off (-12 V shutdown of 3 PIN regulator) |
| Earth leakage current |
<300 µA |
| Safety / EMC |
TÜV (IEC/EN60601-1 3rd Edition), CE |
| Operating temperature |
-10…+70 °C |
| Derating |
2,5 % / K from +50…+70 °C |
| Storage temperature |
-20…+70 °C |
| Operating humidity |
5…95 % RH, non-condensing |
| Dimensions (WxDxH) |
81 x 128 x 40 mm ±0.5 mm |
| Weight (net) |
0.3 kg |
| Product-specific data |
|---|
| Fan use |
8.6 CFM, 15 m3/h |
|
Article |
Output |
Output current |
Output current |
Output current |
Load |
Line |
Ripple & |
|---|---|---|---|---|---|---|---|
| BEP-506M |
+3.3 V |
0 A |
6 A |
4 A |
±4 % |
±1 % at rated load |
50 mV |
| +5 V |
1 A |
8 A |
4 A |
±3 % |
50 mV |
||
| +12 V |
0.2 A |
3 A |
1.5 A |
±4 % |
120 mV |
||
| -12 V |
0 A |
0.5 A |
0.3 A |
±5 % |
120 mV |
||
| +5 Vsb |
0 A |
0.75 A |
±4 % |
120 mV |
Output power is 60 W without fan and 80 W with an 8.6 CFM fan (15 m3/h). Combined output power at +3.3 V and +5 V must not exceed 40 W. The
+5 V output can supply up to 8 A if a minimum load of + 12 V / 0.4 A is connected. The +3.3 V output can supply up to 6 A if a minimum load of + 5 V / 1.2 A and + 12 V / 0.1 A is connected. Ripple and Noise was measured by a 20 MHz bandwidth limited oscilloscope with connected 10 µF electrolytic capacitor and 0.47 µF ceramic capacitor at each output. Load regulation was done by changing the measured output load ±40 % from 60 % rated load and all other outputs kept at 60 % rated load.
As a power component this PSU is for assembly purposes only and must not be operated in unassembled condition. The final assembly has to comply with the valid EMC and safety standards.
| Optional accessories | ||
Article-No. |
Designation |
|
|---|---|---|
| CB-506 | ATX cable harness for BEP-506 | link to the product |
| X1-024 | AC input cable harness / 2 pole / Length: 620 mm / AWG 18 / Ends open | link to the product |