MHG2-6300P-B2

300 Watt

  • Low noise <25 dB(A)!
  • Complies with IEC/EN60601-1 – 3rd Edition
  • Operating temperature -10…+50 °C

The PC power supply MHG2-6300P-B2 is designed for medical applications and in keeping with the IEC/EN60601-1 and UL60601-1 standards. By its integrated temperature-controlled fan regulation it is quiet and, thus, ideal for medical practices and in surgeries. The MHG2-6300P-B2 provides an external grounding bolt, by which the PC can be kept at the same potential capacity as other components when required.

Technical data

 

Input voltage

90…264 VAC, active PFC

Input frequency

47…63 Hz

Input current

7 A (115 VAC)/3,5 A (230 VAC)

Inrush current

65 A (115 VAC) /125 A (264 VAC)

Efficiency

≥75 %, 230 VAC / ≥70 %, 115 VAC (full load)

Hold up time

>16 msec

Power-Good-Signal

Switch on delay 100…500 msec
Switch off delay 1 msec

Protection

Short circuit protection: at each output, switch off / +5 Vsb, auto-recovery
Overload protection: 110…150 %, switch off
Overvoltage protection: +3.3 V (+3.9…+4.3 V), +5 V (+5.7…+6.5 V), +12 V (+13.6…+15 V)

Insulation voltage

Input / Chassis 3100 VDC
Input / Output 4242 VDC

Earth leakage current

<300 µA, 115 VAC/230 VAC

Safety / EMC

TÜV (IEC/EN60601-1 3rd Edition), UL (UL60601-1), CE

Operating temperature

-10…+50 °C

Storage temperature

-40…+80 °C

Operating humidity

20…80 % RH, non-condensing

Loudness

0…200 W / 24 dB (A), 300 W / 26.5 dB (A)

Dimensions

150 x 140 x 86 mm ±0.5 mm

Weight (net)

1.95 kg

Article
No.

Output
voltage

Output current
min

Output current
max

Load
regulation

Ripple &
Noise

MHG2-6300P-B2

+3.3 V

0 A

28 A

±5 %

50 mV

+5 V

1 A

35 A

±5 %

50 mV

+12 V

1 A

22 A

±5 %

120 mV

-12 V

0 A

0.8 A

±5 %

150 mV

-5 V

0 A

0.5 A

±5 %

150 mV

+5 Vsb

0.1A

2 A

±5 %

50 mV

Max. output power is 300 W, combined output current at +3.3 V and +5 V must not exceed 45 A. Ripple and Noise was measured by a 20 MHz bandwidth limited oscilloscope with connected 220 µF electrolytic capacitor and 0.1 µF ceramic capacitor at each output. During a cross regulation test we recommend to keep the channel with higher output load at 80 % of its max. power and the channel with lower output load at 20 % of its max. power.
As a power component this PSU is for assembly purposes only and must not be operated in unassembled condition. The final assembly has to comply with the valid EMC and safety standards.

Technical documentation: