TRG21
20 Watt
- For industrial and medical applications
- High reliabilty (MTBF)
- Compliant with ErP and CEC level V
The series TRG21 desktop adapters supply power to professional devices in commerce, industry and medical applications. Robustly constructed, the adapters are also equipped with top quality components for many years of operation. They comply with current requirements for ecological design and have an idle consumption of less than 0.3 Watt. The entire closed plastic case can easily be desinfected.
| Technical data |
|
|---|---|
| Input voltage |
90…264 V AC |
| Input frequency |
47…63 Hz |
| Inrush current |
40 A max / 240 V AC |
| Efficiency |
App. 75…85 % |
| Idle consumption |
<0.3 W |
| Hold up time |
App. 8 ms / 115 VAC |
| Protection |
Short circuit protection: Switch off with auto recovery |
| Line regulation |
±1 % at rated load and input voltage change of 100…240 VAC |
| Load regulation |
±≤5 % (Load step ±40 % related to 60 % load) |
| Leakage current |
<0.1 mA |
| Safety / EMC |
TÜV (IEC/EN60601-1 3rd Edition), UL (UL60601-1), CAN/CSA-C22.2 No 601.1-M90, TÜV (IEC/EN60950-1), CB, CE |
| Operating temperature |
0…+60 °C |
| Derating |
In the range of +40…+60 °C, 2 % / °C |
| MTBF |
App. 400 000 h according to MIL-HDBK-217F at +25 °C |
| Storage temperature |
-20…+85 °C |
| Operating humidity |
10…93 % RH, non-condensing |
| Dimensions (WxDxH) |
45.2 x 76.2 x 22.1 mm ±0.5 mm |
| Weight (net) |
0.14 kg |
| Product-specific data |
|---|
| ECO standard |
ErP directive 2009/125/EG, CEC level V |
| Max. operation height |
3000 m |
|
Article |
Output |
Output current |
Output current |
Ripple & |
Efficiency |
|---|---|---|---|---|---|
| TRG21A120-CECV |
+12 V |
0 A |
1.8 A |
<120 mVss |
App. 81 % |
| TRG21A180-CECV |
+18 V |
0 A |
1.2 A |
<180 mVss |
App. 84 % |
| TRG21A240-CECV |
+24 V |
0 A |
0.9 A |
<240 mVss |
App. 85 % |
Additional output voltages (+5 V, +9 V, +15 V) are available on request.
Ripple and Noise was measured by a 20 MHz bandwidth limited oscilloscope with connected 100 nF and 10 μF capacitors at the output. Output voltage was adjusted to ±2 % at 60 % load.
The final assembly has to comply with the valid EMC and safety standards.
| Optional accessories | ||
Article-No. |
Designation |
|
|---|---|---|
| KP-01 | KP-01, cling pad, attachment help for easy and simple attachment | link to the product |