MP1S-6400V-B1
400 Watt
- Additional +24 VDC output
- Ideal for applications with large monitors
- High efficiency of up to 84 %
The MP1S-6400V-B1 is a medically approved 1U ATX PC power supply with an additional +24 VDC output. This allows the direct supply of medical peripherals including the areas of diagnostics, therapy and imaging with +24 VDC. The simultaneous supply of motherboard and peripherals with a single power supply also reduces the total earth leakage current of the medical system.
| Technical data |
|
|---|---|
| Input voltage |
100…240 V AC, active PFC |
| Input frequency |
47…63 Hz |
| Input current |
6 A (115 V AC) / 3 A (230 V AC) |
| Inrush current |
70 A (115 V AC / 25°C) / 140 A (230 V AC / 25°C), cold start |
| Efficiency |
84 % (at nominal load and 230 V AC) |
| Hold up time |
>16 ms |
| Power-Good-Signal |
Switch on delay 100…500 ms |
| Protection |
Short circuit protection: +5 V, +12 V, switch off /+3.3 V, -12 V, +5 Vsb, auto recovery |
| Earth leakage current |
<0.3 mA, 115 V AC / 230 V AC |
| Safety / EMC |
TÜV (IEC/EN60601-1 3rd Edition), UL (UL60601-1), CB, CCC, CE |
| Operating temperature |
0…+40 °C |
| MTBF |
149 868 h according to MIL-HDBK-217F at +25 °C with fan |
| Storage temperature |
-20…+80 °C |
| Operating humidity |
20…80 % RH, non-condensing |
| Dimensions (WxDxH) |
81.5 x 250 x 40.3 mm ±0.5 mm |
| Weight (net) |
1.2 kg |
|
Article |
Output |
Output current |
Output current |
Load |
Ripple & |
|---|---|---|---|---|---|
| MP1S-6400V-B1 |
+24 V |
0.1 A |
6 A |
±5 % |
240 mV |
| +3.3 V |
0.5 A |
10 A |
±5 % |
60 mV |
|
| +5 V |
1 A |
12 A |
±5 % |
60 mV |
|
| +12 V |
2 A |
16 A |
±5 % |
120 mV |
|
| -12 V |
0 A |
0.3 A |
±10 % |
120 mV |
|
| +5 Vsb |
0.1 A |
2.5 A |
±5 % |
60 mV |
Max. output power is 400 W. Power at +3.3 V, +5 V, +12 V, –12 V and +5 Vsb is max. 260 W, combined at +3.3 V and +5 V it is 90 W. Ripple and Noise was measured by a 20 MHz bandwidth limited oscilloscope with connected 100 μF electrolytic capacitor and 0.1 μF ceramic capacitor at each output.
As a power component this PSU is for assembly purposes only and must not be operated in unassembled condition. The final assembly has to comply with the valid EMC and safety standards.